Influence of formulation variables and inhalation device on the deposition profiles of cromolyn sodium dry powder aerosols
Kambiz Gilani, Abdolhossien Rouholamini Najafabadi, Majid Darabi, Mohammadali Barghi, Morteza Rafiee-Tehrani

Abstract


Dry powder inhalers (DPIs) have attained considerable attention due to their propellant-free formulations and the patient’s inherent coordination with actuation. Generally, DPI formulations consist of a micronized drug alone or mixed with carrier particles. This study was carried out to investigate the effects of carrier particle size and weight fraction on aerosolisation behaviour of cromolyn sodium (CS).
Pharmatose® 450M and Pharmatose® 325M, two commercial α-lactose monohydrate with different particle sizes, were blended in two different fractions (30 and 50% w/w) with CS. A low resistance device (Spinhaler®) and a medium resistance device (Cyclohaler®), were used to evaluate the effect of inhaler design on the deposition profiles of CS.
The in vitro deposition of the formulations was determined using a twin stage impinger (TSI). Fine particle dose, fine particle fraction and emitted dose of the drug were depended to both formulation and inhalation devices. Fine particle fractions of the drug aerosolised from the formulations ranged from 9.35 up to 36.45%. The highest fine particle fraction was produced by formulation containing 50% Pharmatose® 450M as carrier. Cyclohaler® showed higher efficiency in aerosolisation of CS compared to Spinhaler®.

Keywords


Dry powder inhaler, Cromolyn sodium, Lactose, Particle size, Weight fraction ratio,